Specifications |
PLV®-102
Front Panel Controls
| Mode Selection: |
Control, Control + Sigh, Assist/Control, Assist/Control + Sigh, SIMV |
| Tidal Volume: |
0.05 to 0.20 ± 0.02 liters; 0.20 to 3.00 liters ± 10% |
| Sign Volume: |
150% of Set Tidal Volume
|
| Sign Frequency: |
1 Sign Breath every 100 Tidal
|
| Breath Rate: |
2 to 35 BPM ± 0.5; 36 to 40 BPM ± 2 |
| Inspiratory Flow Rate*: |
10 to 120 LPM |
| Battery Test Switch: |
Test Internal/External Battery |
| Main Power Switch: |
On/Off, Recharge |
| DC Circuit Breaker: |
Push/Pull Actuation |
| AC Circuit Breaker: |
Push/Pull Actuation |
| Airway Pressure Limit: |
5 to 100 cm H2O ± 5 |
| Sensitivity: |
+18 cm H2O ± 1 to at least -6 cm H2O |
| Low Pressure Alarm: |
2 to 40 cm H2O ± 2; 41 to 50 cm H2O ± 5 |
| Manual Sigh: |
Push to Activate One Breath |
| Oxygen Percent**: |
21% to 90% ± 6% |
| Oxygen Sensor (SIMV Operation)**: |
21% to 90% ± 10% |
| 30 Second Alarm Silence:
|
Push to Activate |
* Peak flow rate during inspiratory time
**See Operators Manual for Oxygen Delivery Limitations.
Front Panel Visual Indicators
| Manual Sigh: |
Yellow Indicator Light |
| Tidal Volume: |
Digital Display |
| Rate: |
Patient Breaths Per Minute: Digital Display |
| I:E Ratio: |
Digital Display |
| InspiratoryFlow Rate:
|
Digital Display |
| IncreaseInspiratory Flow: |
Red Indicator Light |
| 120 VAC Power Source: |
Green Indicator Light |
| Internal DC Power Source: |
Amber Indicator Light |
| External DC Power Source: |
Yellow Indicator Light |
| Patient Airway Pressure*: |
-10 to 100 cm H20 ±5 |
| Assisted/Spontaneous: |
Green Indicator Light |
| 15-Second Delay (Low Pressure Alarm): |
Green Indicator Light |
| Oxygen Percent: |
50% and Greater Marked Red |
| 30-Second Alarm Silence: |
Yellow Indicator Light/Push Button |
| Alarm Code Display(in Flow Window): |
Digital Display |
* Pressure waveform is sinusoidal
Alarms
| Low Pressure: |
Visual/Continuous Audible |
| High Pressure: |
Intermittent Audible |
| Apnea: |
Visual/Continuous Audible |
| Inverse I:E Ratio: |
Visual Warning |
| Increase Inspiratory Flow:
|
Visual Warning |
| Low Internal Battery: |
Visual/Continuous Audible |
| Low External Battery:
|
Visual/Continuous Audible |
| Reverse Battery Connection: |
Continuous Audible |
| Power Failure: |
Continuous Audible |
| Unit Malfunction: |
Intermittent Audible |
| Power Source Change: |
Continuous Audible |
| Oxygen: |
Continuous Audible |
Power Sources
- 120 VAC 50/60 Hz
- 100/220/240 VAC 50/60 Hz (International Models)
- 12 VDC External Battery
- 12 VDC Internal Battery
Physical Dimensions
| Overall Dimensions: |
22.9 cm x 31.1 cm x 31.1 cm (9" H x 12.25" W x 12.25" D) |
| Weight: |
13.1 kg (28.9 lb) |
Patents: #4493614, #4617637, #4587967
|